| FAQ: Leica CV5030 Coverslipper | ||||
| Nozzle | ||||
| 01. | Mounting medium is not spread evenly or is applied inconsistently – what is the problem? | |||
| A: The nozzle is clogged with dried medium or dried medium has built up around the nozzle. This can happen, when the dispenser module is left dry for too long and/or if there is a compatibility problem between old and new medium read more ... |
||||
| 02. | The nozzle is dragging the medium over the entire length of the slide (incl. the label area) – what is wrong? | |||
| A: The distance between nozzle and slide is not correct The nozzle needs to be re-adjusted. The recommended distance between nozzle tip and slide is 8 mm. The dispenser module sits on a large screw which can be turned to adjust read more ... |
||||
| 03. | The nozzle keeps falling off – what is the problem? | |||
| A: The nozzle is not securely fastened in the nozzle holder Secure the nozzle correctly in the holder. |
||||
| 04. | The nozzle is clogged with dried medium – what can be done to avoid this problem? | |||
| A:After work and during longer breaks, the nozzle must be parked in the prime position If no coverslipping will be done for more than 1 hour, we recommend placing the dispenser module into the prime position to prevent the nozz read more ... |
||||
|
|
||||