What can I do to avoid noise during sectioning (knife produces a screeching noise) and/or sections with chatter marks or scratches?
FAQ Report
  1. The sectioning speed is too high: Please lower the sectioning speed.
  2. The sectioning thickness is too high: Please lower the thickness.
  3. The clearance angle is too wide: Please try systematically to reduce the clearance angle settings until an optimum clearance angle has been found.
  4. The clamping of the specimen or knife holder might be insufficient: Please check all screws and clamping mechanisms at the specimen holder system and knife holder and retighten, if necessary.
  5. The type of knife is inappropriate: Try and use a more stable knife (e.g. d-profile knife)
  6. The material to be sectioned might be too hard: Please soften the block by soaking it in water