THUNDER Imager Cell Spinning Disk

Clarity Through Synergy

Technical Specifications

Imaging modesWidefield (inc. THUNDER) and Confocal Spinning Disk
Spinning disk unitCrestOptics CICERO
Pin hole size/pin hole spacing50/250 µm
Disk rotation speed15,000 rpm (24 sectors)
WorkflowNavigator
StandDMi8
FocusMotorized, closed-loop, AFC (hardware focus for fast focusing and drift correction)
StageQUANTUM high speed stage (real-time controlled with Synapse)
Transmitted light contrast & methodBF, DIC, Phase Contrast
Emission filter wheel (5-position)
  • 440/40 nm
  • 525/50 nm
  • 600/50 nm
  • Quad D/F/T/Cy5
  • (Empty)
Laser source89-North LDI4
Laser typeMultimode diode lasers
Laser lines
  • 405 nm 300 mW
  • 470 nm 1000 mW
  • 555 nm 1000 mW
  • 640 nm 400 mW
Laser safetyInstalled system is class 1 laser safe
Detection
  • Hammamatsu Fusion
  • Kintetix22
  • (22mm FOV)
IncubationOptional - cage & stage top e.g. Okolab, Tokai Hit
Optional or enhanced image clarity & AI-powered analysisTHUNDER (2D/3D) opto-digital technology, THUNDER Live, Aivia
Optional upgrade path for advanced imaging workflows
  • Adaptive Immersion (40 x and 63 x adaptive immersion objectives)
  • SmartCORR
  • Sample Finder*
  • Enhanced Navigator

*Full Sample Finder functionality requires scanning stage

Scroll to top