STELLARIS 5 Cryo Confocal Light Microscope

Approval

Conformity Marking

CE (European Union)

Electrical

Mains Voltage

100/115/230 V (+/- 10%), 50/60 Hz

Power Consumption

~ 100

Control Voltage

24 V DC

Dimensions and Weight

Width x Depth x Height

Weight

Cover Box

340 x 320 x 305 mm

~ 10 KG

Cryo Stage

243 x 142 x 77 mm

~ 1 KG

Cryo Transfer Shuttle

393 x 120 x 119 mm

~ 1 KG

LN2 Pump Controller

220 x 354 x 132 mm

~ 5 KG

LN2 Pump

1170 x 92 x 92 mm

~ 2 KG

LN2 Dewar (Empty)

200 x 200 x 758 mm

~ 10 KG

LN2 Dewar (Full)

~ 30 KG

Operating Conditions

Min/Max

Optimal optical behavior

Room Temperature

18 to 25 °C +/- 1 °C

22 °C +/- 0,5 °C

Room Rel. Humidity

< 40%

< 20%

Storage Temperature

5 to 35 °C

21 to 24 °C

Storage Humidity

Max. 50%

Around 40%

Room Ventilation

Required at all times during operation

Operating Temperatures

Cooling Temperature

-195 °C

Heating Range

Up to 60 °C

Cooling Time

20 Minutes (From +22 to -195 °C)

Cryo Stage Metal Shell Range

5 to 45 °C (Avoids Icing)

Dewar

LN2 Dewar Volume

25 l

Dewar Temperature

-196 °C